Heraeus FiberLight D2 modules are available in different versions. They come with standard or extended wavelength range and either focused (with SMA 905 connector) or parallel beam outlet. Summarized, FiberLight modules from Heraeus Nobelight are characterized by the following advantages:
· Compact size · Low power consumption (6 W) · Low heat dissipation · Easy integration of optical fibers, measuring cells and capillaries |
· Instant lamp ignition · Cyclic operation · Extended service life of up to 3 years · Shutter function |
TECHNICAL DATA AND SPECIFICATIONS
FiberLight Module |
 |
|
FiberLight D2 Basic |
|
Spectral distribution: |
200-1100 nm, optional 185-1100 nm |
|
Power consumption: |
6 W |
|
Power requirements: |
12 Vdc/0.6 Adc |
|
Operating ambient temperature: |
5-35° C |
|
Relative humidity: |
max. 90%, non condensing |
|
Light exit: |
focused* or collimated** |
|
Optical fiber Ø: |
200, 400 and 600 µm* |
|
Optical fiber connector: |
SMA 905* |
|
Numerical aperture NA: |
D2-lamp 0.245; W-lamp 0.057 |
|
Cooling: |
not required |
|
Weight: |
130 g |
|
Dimensions (L x W x H): |
157 x 55 x 37 mm |
|
External control: |
on/off D2- and W-lamp (independently) shutter open/close |
|
Deuterium Lamp D2 Basic |
Spectral distribution: |
200-400 nm line free, optional 185-400 nm*** |
|
Window material: |
fused quartz, fused synthetic silica*** |
|
Light output (radiant intensity): |
≥5x10-8 W/Sr @ 240 nm |
|
Stability: |
≤1x10-3 AU |
|
Drift: |
≤0.25%/h |
|
Excitation frequency: |
250 kHz |
|
Ignition voltage: |
approx. 1 kV |
|
Life: |
≥1000 h @ 230 nm (50% intensity loss) |
|
Tungsten Lamp |
Spectral distribution: |
400-1100 nm |
|
Voltage: |
5 Vdc |
|
Current: |
45 mAdc |
|
Typical lifetime: |
≥2000 h |
|
|
* FiberLight D2 Basic DTM 6/10, DTM 6/10S, DTM 6/15, DTM 6/50, DTM 6/50S ** FiberLight D2 Basic DTM 6/11, DTM 6/11S *** FiberLight D2 Basic DTM 6/10S, DTM 6/11S, DTM 6/50S |
|
|
CYCLIC OPERATION AND LIFETIME The continuous operating life of a FiberLight D2 deuterium lamp exceeds 1000 hours. As electrodeless discharge lamps (EDL) with high frequency excitation come into use, FiberLight D2 can be switched on and off on demand and operated in cycles. Due to the fact that the number of ignitions does not have a negative effect on the lamp's lifetime, an extended service life of up to three years can be achieved. Additionally pulse-to-pulse repeatability is extremely consistent to within 0,1%.
INSTANT ON AND STABILITY The FiberLight D2 EDL ignites instantly and therefore delivers an immediate stable light output. This feature makes it unique among UV light sources. FiberLight D2 is therefore the ideal light source in analytical instrumentation for waste water analysis and other pollution monitoring, where light absorption is measured for only a few seconds and repeated after long intervals. Consequently the lamp would be switched on only for the short measurement time, while it would remain turned off for most of the time.
APPLICATION EXAMPLE To determine the nitrate content in waste water, FiberLight D2 is switched on for the duration of ten seconds to measure the nitrate light absorption. This procedure is repeated every 300 seconds. Measurement consistency is extremely good and the lamp's lifetime can be extended up to three years under these operating conditions.
Bottom image: FiberLight D2 DTL lamp unit, consisting of D2- and W-lamp
Top right image: Typical lifetime FiberLight D2 (230 nm)
 |
Center right image: Cyclic operation at 230 nm (measuring time: 10 s on / 300 s off)
Bottom right image: Spectral distribution, D2 Basic (D2 HighPower not available anymore) |
|
   |
ORDER INFORMATION
Description |
Part No. |
Price |
FiberLight Modules D2 |
FiberLight D2 Basic DTM 6/10 - 1,0 mm aperture, quartz glass, focused, for 400-600 µm fibers, 200-1100 nm |
80007444 |
835,00 |
|
FiberLight D2 Basic DTM 6/50 - 0,5 mm aperture, quartz glass, focused, for 200-600 µm fibers, 200-1100 nm |
45006281 |
835,00 |
|
FiberLight D2 Basic DTM 6/10S - 1,0 mm aperture, syn. quartz, focused, for 400-600 µm fibers, 185-1100 nm |
80000755 |
835,00 |
|
FiberLight D2 Basic DTM 6/50S - 0,5 mm aperture, syn. quartz, focused, for 200-600 µm fibers, 185-1100 nm |
80001017 |
835,00 |
|
FiberLight D2 Basic DTM 6/11 - 1,0 mm aperture, quartz glass, collimated beam (quasi parallel light) |
80003030 |
835,00 |
|
FiberLight D2 Basic DTM 6/11S - 1,0 mm aperture, syn. quartz, collimated beam (quasi parallel light) |
80048581 |
835,00 |
|
FiberLight D2 Basic DTM 6/15 - 1,0 mm aperture, quartz glass, focused, for 400-600 µm fibers, 200-1100 nm |
80010770 |
835,00 |
FiberLight Modules UV-LED |
FiberLight L3 UV-LED |
80158546 |
941,00 |
FiberLight Replacement Lamps |
FiberLight D2 Replacement Lamp DTL 6/10 - for DTM 6/10 |
45006253 |
452,00 |
|
FiberLight D2 Replacement Lamp DTL 6/50 - for DTM 6/50 |
45006266 |
452,00 |
|
FiberLight D2 Replacement Lamp DTL 6/10S - for DTM 6/10S |
80000756 |
452,00 |
|
FiberLight D2 Replacement Lamp DTL 6/50S - for DTM 6/50S |
80001018 |
452,00 |
|
FiberLight D2 Replacement Lamp DTL 6/15 - for DTM 6/15 |
80010771 |
452,00 |
|
FiberLight D2 Replacement Lamp DTL 10/50S - for DTM 10/50S |
80088565 |
612,00 |
|
|
All prices are in Euro (EUR) per piece (excluding VAT) and are subject to change without further notice.
MORE INFORMATION
Heraeus brochure - "FiberLight Miniature UV-Vis Light Source" |
PDF: 1,9 MB |
|
FiberLight application note - "Measuring and Analyzing Seawater" |
PDF: 473 KB |
|
FiberLight application note - "Measuring and Analyzing Coastal
Sea Contamination" |
PDF: 391 KB |
|
FiberLight application note - "Beer-Lambert Law - All about the beer! - Spectroscopic Quality Control" |
PDF: 363 KB |
|
|
All brand names and registered trademarks are the property of their respective owners. All images copyright protected. Subject to modifications and errors.
For further questions, please don't hesitate to contact us. |